Understanding how to care for the OneTouch Verio Reflect® meter.
Product Guide
Global
OneTouch® Delica® Plus lancets; OneTouch® Delica® Plus lancing device; OneTouch Verio® control solution; OneTouch Verio Reflect® meter; OneTouch Verio® test strips
English
0.0
Article Details
Caring for Your System:

Storing your system:
  • Store your meter, test strips, control solution and other items in your carrying case. Keep in a cool, dry place between 5°C and 30°C. Do Not refrigerate. Keep all items away from direct sunlight and heat. Exposure to temperatures outside the storage limits, as well as moisture and humidity, may cause inaccurate readings.
Cleaning and disinfection:
  • The meter, lancing device and cap should be cleaned whenever they are visibly dirty and before disinfection. Clean your meter at least once per week. For cleaning obtain regular strength liquid dish soap and a soft cloth. Prepare a mild detergent solution by stirring 2.5 mL of regular strength liquid dish soap into 250 mL of water.
    • Do Not use alcohol or any other solvent.
    • Do Not allow liquids, dirt, dust, blood or control solution to enter the test strip port or the data port.
    • Do Not spray cleaning solution on the meter or immerse it in any liquid.
1. Holding the meter with the test strip port pointed down, use a soft cloth dampened with water and mild detergent to wipe the outside of the meter and lancing device
image.png
  • Be sure to squeeze out any excess liquid before you wipe the meter. Wipe the outside of the cap.
2. Wipe dry with a clean, soft cloth

 
  • The meter, lancing device and cap should be disinfected periodically. Clean your meter, lancing device and cap prior to disinfecting. For disinfecting, obtain regular household bleach (containing a minimum of 5.5% sodium hypochlorite as the active ingredient)*. Prepare a solution of 1 part household bleach and 9 parts water or obtain equivalent (0.55% sodium hypochlorite*) wipes.
    • ​​​​​​​*Follow manufacturer's instruction for handling and storage of bleach.
1. Hold the meter with the test strip port pointed down
image.png
  • Use a soft cloth dampened with this solution or a 0.55% equivalent sodium hypochlorite wipe to wipe the outside of the meter and lancing device until the surface is damp. Be sure to squeeze out any excess liquid before you wipe the meter.
2. After wiping, cover the surface you are disinfecting with the soft cloth dampened with the bleach solution or 0.55% equivalent sodium hypochlorite wipe for 1 minute
image.png
  • Then wipe with a clean, damp, soft cloth.
  • Wash hands thoroughly with soap and water after handling the meter, lancing device and cap.
  • If you see signs of wear, please contact Customer Care. 