Understanding error and other messages that may appear on the OneTouch Verio Reflect® meter such as Extreme Low Glucose, Extreme High Glucose and temperature too high or too low.
Product Guide
Global
OneTouch Verio Reflect® meter
English
0.0
Article Details
Error and other messages:
 
  • The OneTouch Verio Reflect® Meter displays messages when there are problems with the test strip, with the meter or when your glucose levels are above 33.3 mmol/L or below 1.1 mmol/L. Improper use may cause an inaccurate result without producing an error message.
  • NOTE:
    • If the meter is on but does not operate (locks up), contact Customer Care.
Extreme Low Glucose:
image.png
  • What it means:
    • You may have a very low blood glucose level (severe hypoglycemia).
  • What to do:
    • This may require immediate treatment. Although this message could be due to a test error, it is safer to treat first and then do another test.
    • Always treat according to your healthcare professional's recommendations.
Extreme High Glucose:
image.png
  • What it means:
    • You may have a very high blood glucose level (severe hyperglycemia), over 33.3 mmol/L.
  • What to do:
    • Re-test your blood glucose level. If the result is EXTREME HIGH GLUCOSE again, obtain and follow instructions from your healthcare professional right away.
Temperature too high:
image.png
  • What it means:
    • Meter is too hot (above 44°C) to perform a test.
  • What to do:
    • Move the meter and test strips to a cooler area. Insert a new test strip when the meter and test strips are within the operating range (6-44°C). If you do not get another Temperature too high message, you can proceed with testing.
    • If this message continues to appear, contact Customer Care.
Temperature too low:
  • What it means:
    • Meter is too cold (below 6°C) to perform a test.
  • What to do:
    • Move the meter and test strips to a warmer area. Insert a new test strip when the meter and test strips are within the operating range (6-44°C).
    • If you do not get another Temperature too low message, you can proceed with testing.
    • If this message continues to appear, contact Customer Care. 