Salesforce

OneTouch® Ultra®2 – When To Test With Control Solution

« Go Back
Information
OneTouch® Ultra®2 – When To Test With Control Solution
OT-Ultra-2-When-to-Test-with-Control-Solution
Information for when to test with control solution when using the OneTouch® Ultra®2 meter.
Product Guide
Global
OneTouch® Ultra®2 meter; OneTouch Ultra® control solution
English
3.3
Article Details
When to Test with Control Solution:
 
  • OneTouch Ultra® Control Solution contains a known amount of glucose and is used to check that the meter and the test strips are working properly.
  • Do a control solution test:
    • whenever you open a new vial of test strips.
    • if you suspect the meter and test strips are not working properly.
    • if you have had repeated unexpected blood glucose results.
    • if you drop or damage the meter.
  • Note:
    • Use only OneTouch Ultra® Control Solution with your OneTouch® Ultra® 2 Meter.
    • Control solution tests must be done at room temperature (68–77°F). Make sure your meter, test strips, and control solution are at room temperature before testing.
    • Write the discard date (3 months after first opening the vial) on the vial label when you first open it.
  • CAUTION: Do Not swallow control solution; it is not for human consumption. Do Not apply control solution to the skin or eyes as it may cause irritation.