Error 2:
 
image.png
  • What it means:
    • Both Error 2.1 and 2.2 can be caused by either:
      • Use of a used test strip.
      • Damaged test strip. Check the bottom of the strip.
      • Meter contamination such as control solution or blood have gone inside the test strip port.
  • What to do:
    • Check test strip port for contamination. Confirm if any control solution or blood have gone inside the test strip port.
    • Repeat the test with a new test strip.
    • If this message continues to appear, contact Customer Care.