OneTouch® Select Simple™ - System Accuracy

« Go Back
Information
OneTouch® Select Simple™ - System Accuracy
OneTouch-Select-Simple-System-Accuracy-india
System Accuracy of your OneTouch® Select Simple™ meter.
Product Guide
Global
OneTouch Select Simple® meter
English
0.0
Article Details

System Accuracy:

 

  • Diabetes experts have suggested that glucose meters should agree within ±15 mg/dL of a laboratory method when the glucose concentration is lower than 75 mg/dL, and within 20% of a laboratory method when the glucose concentration is 75 mg/dL or higher. Samples from 100 diabetic patients at 1 clinical centre were tested using both the OneTouch® SelectSimple™ System and the YSI Model 2300 Glucose Analyser (laboratory test).

    • Therefore, 99.7% of the total results obtained with the OneTouch® SelectSimple™ System achieved the goal suggested by the diabetes experts.

 

Regression Statistics:

  • Samples were tested in duplicate on each of three test strip lots. Results indicate that the OneTouch® SelectSimple™ Meter compares well with a laboratory method.

AW 06693803 A