Understanding a Warning Temp Too Low error message on the OneTouch Verio IQ® meter.
Product Guide
Global
OneTouch VerioIQ® meter
English
0.0
Article Details
Troubleshooting:
  • The OneTouch Verio IQ® Meter displays messages when there are problems with the test strip, with the meter, or when your glucose levels are above 600 mg/dL or below 20 mg/dL. Improper use may cause an inaccurate result without producing an error message.
  • Note: If the meter is on but does not operate (locks up), contact OneTouch® Customer Care.

Temperature Too Low:
  • What it means:
    • Meter is too cold (below 43°F) to work correctly.
  • What to do:
    • Move the meter and test strips to a warmer area. Insert a new test strip when the meter and test strips are within the operating range (43–111°F). If you do not get another Temperature too low message, you can proceed with testing.
AW 06703306A