Understanding a Warning Temp Too Low error message on the OneTouch Verio IQ® meter.
Product Guide
Global
OneTouch VerioIQ® meter
English
0.0
Article Details
Troubleshooting:
The OneTouch Verio IQ®Meter displays messages when there are problems with the test strip, with the meter, or when your glucose levels are above 600 mg/dL or below 20 mg/dL. Improper use may cause an inaccurate result without producing an error message.
Note: If the meter is on but does not operate (locks up), contact OneTouch®Customer Care.
Temperature Too Low:
What it means:
Meter is too cold (below 43°F) to work correctly.
What to do:
Move the meter and test strips to a warmer area. Insert a new test strip when the meter and test strips are within the operating range (43–111°F). If you do not get another Temperature too low message, you can proceed with testing.