Salesforce

OneTouch Verio Reflect® - Testing with Control Solution

« Go Back
Information
OneTouch Verio Reflect® - Testing with Control Solution
OneTouch-Verio-Reflect-Testing-with-Control-Solution-canada
How to perform a control solution test, view control solution results and check if the result is in range with the OneTouch Verio Reflect® meter as well as instructions on how to clean the control solution cap.
Product Guide
Global
OneTouch Verio® control solution; OneTouch Verio Reflect® meter; OneTouch Verio® test strips
English
0.0
Article Details
Testing with control solution:

Control solution testing precautions:
  • OneTouch Verio® Control Solution is used to check that the meter and test strips are working together properly and that the test is performing correctly. (Control solution is available separately.)
  • NOTE:
    • Use only OneTouch Verio® Control Solution with your OneTouch Verio Reflect® Meter. 
    • When you first open a new vial of control solution, record the discard date on the vial label. Refer to the control solution insert or vial label for instructions on determining the discard date.
    • Tightly close the cap on the control solution vial immediately after use to avoid contamination or damage.
  • CAUTION:
    • Do Not swallow or ingest control solution.
    • Do Not apply control solution to the skin, eyes, ears or nose as it may cause irritation.
    • Do Not use control solution after the expiration date (printed on the vial label) or the discard date, whichever comes first, or your results may be inaccurate.
When to do a control solution test:
  • When you open a new vial of test strips.
  • If you suspect that the meter or test strips are not working properly.
  • If you have had repeated unexpected blood glucose readings.
  • If you drop or damage the meter.
Preparing your meter for a control solution test:
 
1. Press and hold OK to turn the meter on
  • Wait for the Main Menu to appear.
2. Press ↑ or ↓ to highlight Settings, then press OK.
3. Highlight Control Test, then press OK.
  • Wait for the Insert Strip screen to appear on the display.
4. Insert a test strip into the test strip port
image.png
5. Wait for the Apply Control Solution screen to appear on the display
image.png

Preparing the control solution:
 
1. Before removing the cap, shake the vial gently

2. 
Remove the vial cap and place it on a flat surface with the top of the cap pointing up

3. Squeeze the vial to discard the first drop
image.png
4. Wipe both the tip of the control solution vial and the top of the cap with a clean, damp tissue or cloth
5. Squeeze a drop into the small well on the top of the cap or onto another clean, non-absorbent surface
image.png

Applying the control solution:
 
1. Hold the meter so that the side edge of the test strip is at a slight angle to the drop of control solution
image.png
2. Touch the channel on the side of the test strip to the control solution
 
3. Wait for the channel to fill completely
image.png

Viewing your control solution result:
  • After the control solution is applied, your meter will display a progress screen for approximately 5 seconds. Your result is displayed along with the date, time, unit of measure, and (for control solution).
  • Control solution results are stored in the meter and can be seen when reviewing past results on the meter.
  • CAUTION:
    • Make sure you select Control Test from the Settings Menu before you begin a control solution test.
    • If  does not appear on the screen, this result will be included in your averages and your averages will change too. Repeat the test with a new test strip. If the problem persists, contact Customer Care. 
Checking if the result is in range:
  • Compare the result displayed on the meter to the range printed on your OneTouch Verio® Control Solution vial. If your control solution result falls outside the expected range, repeat the test with a new test strip.
  • CAUTION:
    • The range printed on the control solution vial is for control solution tests only and is not a recommended range for your blood glucose level.
Causes of out-of-range results:
  • Out-of-range results may be due to:
    • Not following the instructions for performing a control solution test.
    • Control solution is contaminated, expired, or past its discard date.
    • Test strip or test strip vial is damaged, or past its expiration date.
    • Meter, test strips and/or control solution were not all at the same temperature when the control solution test was performed.
    • A problem with the meter.
    • Dirt or contamination in the small well on the top of the control solution cap.
  • CAUTION:
    • If you continue to get control solution results that fall outside the range printed on the control solution vial, Do Not use the meter, test strips, or control solution. 
    • Contact Customer Care.


Cleaning the control solution cap:

  • Clean the top of the control solution cap with a clean, damp tissue or cloth.