Testing your blood glucose:
Preparing for a blood glucose test:
- NOTE:
- Many people find it helpful to practice testing with control solution before testing with blood for the first time.
Have these things ready when you test:
- OneTouch Verio Reflect® Meter
- OneTouch Verio® Test Strips
- Lancing device
- Sterile lancets
- NOTE:
- Use only OneTouch Verio® Test Strips.
- Unlike some blood glucose meters, no separate step to code your OneTouch Verio Reflect® System is required.
- Testing must be done within operating temperature range (6-44°C).
- Make sure your meter and test strips are about the same temperature before you test.
- Keep test strips in a cool, dry place between 5°C and 30°C.
- Do Not test if there is condensation (water build-up) on your meter. Move your meter and test strips to a cool, dry spot and wait for the meter surface to dry before testing.
- Do Not open the test strip vial until you are ready to remove a test strip and perform a test. Use the test strip immediately after removing it from the vial.
- Tightly close the cap on the vial immediately after use to avoid contamination and damage.
- Store unused test strips only in their original vial.
- Do Not return the used test strip to the vial after performing a test.
- Do Not re-use a test strip that had blood, control solution, or any contaminants applied to it. Test strips are for single use only.
- With clean, dry hands, you may touch the test strip anywhere on its surface. Do Not bend, cut or modify the test strip in any way.
- Comparing your blood glucose test results taken with this meter to your results taken from a different meter is not recommended. Results may differ between meters and are not a useful measure of whether your meter is working properly. To check your meter accuracy, you should periodically compare your meter results to those obtained from a lab.
- IMPORTANT:
- If another person assists you with testing, the meter should always be cleaned prior to use by that person.
- CAUTION:
- The OneTouch Verio Reflect® Blood Glucose Monitoring System should not be used for patients within 24 hours of receiving a D-xylose absorption test as it may cause inaccurately high results.
- Do Not use the OneTouch Verio® Family of Meters when PAM (Pralidoxime) is known or suspected to be in the patient's whole blood sample.
- Do Not use your test strips if your vial is damaged or left open. This could lead to error messages or inaccurate results. Contact Customer Care immediately if the test strip vial is damaged.
- If you cannot test due to a problem with your testing supplies, contact your healthcare professional. Failure to test could delay treatment decisions and lead to a serious medical condition.
- The test strip vial contains drying agents that are harmful if inhaled or swallowed and may cause skin or eye irritation.
- Do Not use test strips after the expiration date (printed on the vial) or your results may be inaccurate.